WEKO3
アイテム
Pellet charge exchange helium measurement using neutral particle analyzer in large helical device
http://hdl.handle.net/10655/3811
http://hdl.handle.net/10655/38115a22e5e9-d62f-4c2f-971f-e2bb14a90ae4
名前 / ファイル | ライセンス | アクション |
---|---|---|
1899 RevSciInstrum_79_10E518.pdf (528.0 kB)
|
Item type | 学術雑誌論文 / Journal Article_02(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2010-03-08 | |||||
タイトル | ||||||
タイトル | Pellet charge exchange helium measurement using neutral particle analyzer in large helical device | |||||
言語 | ||||||
言語 | eng | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | alpha-particles | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | fusion reactors | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | helium | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | hydrogen | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | plasma collision processes | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | plasma | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | confinement | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | plasma diagnostics | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | plasma radiofrequency heating | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | stellarators | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | journal article | |||||
著者 |
Ozaki, T.
× Ozaki, T.× Goncharov, P.× Veshchev, E.× Tamura, N.× Sudo, S.× Seki, T.× Kasahara, H.× Takase, Y.× Ohsako, T. |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | It is very important to investigate the confinement of alpha particles, which will be produced by nuclear reactions in ITER and fusion reactors. The pellet charge exchange (PCX) measurement is one of the most powerful methods because it can directly provide the profile of the alpha particle energy spectra in a plasma. In the large helical device, PCX using tracer encapsulated solid pellet (TESPEL) has been tried in many hydrogen and helium plasmas, including helium accelerated by using the cyclotron resonance heating. In the PCX, we use the compact neutral particle analyzer without simultaneous mass separation ability. The helium particle measurement can be achieved by the application of voltage in the condenser plate. The scattering of hydrogen particle is carefully considered during the estimation of the helium amount. The radial helium profiles can also be obtained by comparing four TESPEL injection shots with/without higher harmonic fast wave heating and at applied plate voltages for He or H, respectively. | |||||
書誌情報 |
en : Review of Scientific Instruments 巻 Vol.79, p. 10E518-1 - 10E518-4, 発行日 2008-10-01 |
|||||
出版者 | ||||||
出版者 | American Institute of Physics | |||||
DOI | ||||||
関連タイプ | isIdenticalTo | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1063/1.2978191 | |||||
権利 | ||||||
権利情報 | ? 2008 American Institute of Physics. | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.1063/1.2978191 | |||||
関連名称 | Publisher version | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |