内容記述タイプ | Abstract | |||||
内容記述 | In situ calibration of the neutron activation system on the Large Helical Device (LHD) was performed by using an intense 252Cf neutron source. To simulate a ring-shaped neutron source, we installed a railway inside the LHD vacuum vessel and made a train loaded with the 252Cf source run along a typical magnetic axis position. Three activation capsules loaded with thirty pieces of indium foils stacked with total mass of approximately 18 g were prepared. Each capsule was irradiated over 15 h while the train was circulating. The activation response coefficient (9.4 ± 1.2) × 10−8 of 115In(n, n′)115mIn reaction obtained from the experiment is in good agreement with results from three-dimensional neutron transport calculations using the Monte Carlo neutron transport simulation code 6. The activation response coefficients of 2.45 MeV birth neutron and secondary 14.1 MeV neutron from deuterium plasma were evaluated from the activation response coefficient obtained in this calibration experiment with results from three-dimensional neutron calculations using the Monte Carlo neutron transport simulation code 6. | |||||
書誌情報 |
Review of Scientific Instruments 巻 88, 号 11, p. 113302, 発行日 2017-11 |
|||||
出版者 | ||||||
出版者 | AIP Publishing | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 00346748 | |||||
書誌レコードID | ||||||
識別子タイプ | NCID | |||||
関連識別子 | 00346748 | |||||
DOI | ||||||
関連タイプ | isIdenticalTo | |||||
識別子タイプ | DOI | |||||
関連識別子 | 10.1063/1.5009475 | |||||
権利 | ||||||
権利情報 | This article may be downloaded for personal use only. Any other use requires prior permission of the author and AIP Publishing. This article appeared in Review of Scientific Instruments 88, 113302 (2017) and may be found at https://aip.scitation.org/doi/abs/10.1063/1.5009475 | |||||
権利 | ||||||
権利情報 | Copyright 2021 Author(s). This article is distributed under a Creative Commons Attribution (CC BY) License. | |||||
情報源 | ||||||
関連名称 | Review of Scientific Instruments 88, 113302 (2017); https://doi.org/10.1063/1.5009475 | |||||
関連サイト | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.1063/1.5009475 | |||||
関連名称 | Publisher version | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
NAIS | ||||||
値 | ||| |